Z-1595-2026 Class II Ongoing

Recalled by I.T.S. GmbH — Lasnitzhohe

Recall Details

Product Type
Devices
Report Date
April 1, 2026
Initiation Date
January 23, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4 units

Product Description

FR.O.H. Calcaneus Repair System with the following description: Anterolateral Calcaneus Plate. Article Number: 21064-60.

Reason for Recall

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Distribution Pattern

US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Code Information

Article Number (Lot Numbers): 21064-60 (679/012318, 679/0722100).