Z-1600-2026 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- April 1, 2026
- Initiation Date
- January 23, 2026
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 11 units
Product Description
Pelvic Reconstruction System (PRS) with the below descriptions and corresponding article numbers. 1. Fixation Screw; Article Number: 70312. 2. Spike short; Article Number: 70314.
Reason for Recall
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Distribution Pattern
US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
Code Information
1. Article Number (Lot Number): 70312 (26/F65133). 2. Article Number (Lot Number): 70314 (86/619736).