Z-1593-2018 Class II Terminated

Recalled by Cook Inc. — Bloomington, IN

FDA device recall Z-1593-2018 was initiated by Cook Inc. on Jan 31, 2018 and is designated Class II. Reason for recall: The interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use. The recall status is terminated (terminated Apr 17, 2020). Affected quantity: 370.

Recall Details

Product Type
Devices
Report Date
May 9, 2018
Initiation Date
Jan 31, 2018
Termination Date
Apr 17, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
370

Product Description

StoneBreaker Probe Cap, Catalog number:SBA-PC; GPN: G52601; UDI:(01)00827002526013

Reason for Recall

The interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.

Distribution Pattern

US Nationwide, OUS countries: AE, AT, AU, BE, BG, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DJ, DK, DO, EC, EE, EG, ES, FI, FJ, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IT, JO, KZ, LB, LT, LV, MA, MO, MX, MY, NI, NL, NO, NZ, PE, PH, PK, PL, PR, PT, QA, RO, RU, SA, SE, SG, SI, SK, SV, TH, TT, TW, UA, UY, XK, ZA

Code Information

All Lots

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1593-2018
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.