Z-1597-2018 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc. — Hoffman Estates, IL

FDA device recall Z-1597-2018 was initiated by Siemens Medical Solutions USA, Inc. on Jan 26, 2018 and is designated Class II. Reason for recall: Error introduced into PET images acquired and reconstructed with VJ20A software. Array values are indexed improperly when the norm file is created during QC. During data reconstruction, incorrect… The recall status is terminated (terminated Oct 12, 2018). Affected quantity: 42 units.

Recall Details

Product Type
Devices
Report Date
May 9, 2018
Initiation Date
Jan 26, 2018
Termination Date
Oct 12, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
42 units

Product Description

Biograph Horizon - PET/CT, PETsyngo VJ20A Software Nuclear medicine/ xray diagnostic scanner.

Reason for Recall

Error introduced into PET images acquired and reconstructed with VJ20A software. Array values are indexed improperly when the norm file is created during QC. During data reconstruction, incorrect values are being applied. This can lead to a gradient in the image. The severity of the error is directly related to the positioning of the PET QC phantom relative to the center of the field of view.

Distribution Pattern

Worldwide distribution. US nationwide, Australia, France, India, Japan, and Netherlands.

Code Information

Software version VJ20A. Code No. 10532746, 10532748.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1597-2018
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.