Z-1601-2024 Class II Ongoing

Recalled by Siemens Medical Solutions USA, Inc. — Issaquah, WA

FDA device recall Z-1601-2024 was initiated by Siemens Medical Solutions USA, Inc. on Feb 23, 2024 and is designated Class II. Reason for recall: On ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into… The recall status is ongoing. Affected quantity: 67.

Recall Details

Product Type
Devices
Report Date
May 1, 2024
Initiation Date
Feb 23, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
67

Product Description

ACUSON Maple Diagnostic Ultrasound System, REF: 11711750, ACUSON Maple Select, REF: 11711779, running software version VA10A

Reason for Recall

On ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into the SR feature, then The SR viewer will display the LAST measured value, not min or max, which could contribute to patient condition misdiagnosis or negatively influence patient management decisions.

Distribution Pattern

Worldwide distribution - US Nationwide including in the states of AL, AR, CA, CO , CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA , MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SD, TN, TX, UT, VA, WA, WI, WY and the countries of BG, BW, BA, TR, AR, CL, CN, FR, SK, CD, BH, PH, AM, CR, AF, BE, AE, AZ, AO, KR, RO, HU, HR, DK, AU, DZ, CY, PL, BO, AL, NL, CI, BR, CZ, DE, AT, ES, CO, BD, CH, IT, CA, CM, DO, EC, EG, ER, ET, FI, FO, GB, GE, GR, GT, GY, HK, HN, ID, IE, IL, IN, IQ, IR, JO, JP, KE, KW, KZ, LB, LT, LU, LV, LY, MA, MD, ME, MK, MM, MN, MU, MW, MX, MY, NG, NO, NP, NZ, OM, PA, PE, PG, PK, PT, PY, QA, RS, RU, SA, SE, SG, SI, SM, SN, SV, SZ, TH, TN, TT, TW, TZ, UA, UG, UY, UZ, VN, XK, YE, ZA, ZM, ZW.

Code Information

REF/UDI-DI: 11711750/04056869995809, and 11711779/04056869995830. Serial Numbers: 700127 700125 700109 700123 700124 700126 700131 700077 700081 700070 700071 700073 700075 700076 700072 700074 700096 700097 700098 700100 700103 700082 700030 700036 700037 700028 700029 700031 700033 700022 700023 700024 700025 700035 700121 700111 700132 700065 700058 700026 700019 700020 900581 900582 900583 900585 700034 700032 700041 700040 700039 700052 700053 700057 700038 700063 700062 700060 700064 700068 700059 700084 700083 700107 700114 700115 700135.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1601-2024
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.