Z-1605-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- May 9, 2018
- Initiation Date
- March 26, 2018
- Termination Date
- June 12, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 37
Product Description
MultiDiagnost Eleva 708036 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.
Reason for Recall
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
Distribution Pattern
Worldwide Distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Farce Islands, Finland, France, Germany, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Korea, Republic of, Latvia, Lebanon, Luxembourg, Malaysia, Martinique Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.
Code Information
System Serial # 18 42 46 41 36 35 37 47 27 43 10 5 24 29 34 49 26 13 15 25 50 12 17 48 32 30 31 14 40 19 9 16 45 8 4 21 11