Z-1607-2018 Class II Terminated
FDA device recall Z-1607-2018 was initiated by Philips Medical Systems Nederlands on Mar 26, 2018 and is designated Class II. Reason for recall: The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot… The recall status is terminated (terminated Jun 12, 2020). Affected quantity: 199.
Recall Details
- Product Type
- Devices
- Report Date
- May 9, 2018
- Initiation Date
- Mar 26, 2018
- Termination Date
- Jun 12, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 199
Product Description
Reason for Recall
Distribution Pattern
Worldwide Distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Farce Islands, Finland, France, Germany, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Korea, Republic of, Latvia, Lebanon, Luxembourg, Malaysia, Martinique Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.