Z-1615-2014 Class II Terminated

Recalled by Biomet, Inc. — Warsaw, IN

FDA device recall Z-1615-2014 was initiated by Biomet, Inc. on Apr 22, 2014 and is designated Class II. Reason for recall: This lot of the ExploR Modular Radial Stem may be missing the wedge ramp threads. If the wedge ramp threads are missing from the stem, the likelihood of the set screw loosening from the radial stem… The recall status is terminated (terminated Jan 13, 2015). Affected quantity: 8 devices.

Recall Details

Product Type
Devices
Report Date
May 21, 2014
Initiation Date
Apr 22, 2014
Termination Date
Jan 13, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8 devices

Product Description

ExploR 7x26mm Modular Radial Stem, Catalog number 11-210062. Elbow prosthesis orthopedic implant.

Reason for Recall

This lot of the ExploR Modular Radial Stem may be missing the wedge ramp threads. If the wedge ramp threads are missing from the stem, the likelihood of the set screw loosening from the radial stem intraoperatively is increased. If the set screw loosens, a revision surgery may be necessary.

Distribution Pattern

US Nationwide Distribution in the states of NY, MD, NC, SC, MI, OR, and NV.

Code Information

Catalog Number: 11-210062 Lot Number Identification: 972230

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1615-2014
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.