Z-1621-2013 Class II Terminated

Recalled by Zimmer, Inc. — Warsaw, IN

FDA device recall Z-1621-2013 was initiated by Zimmer, Inc. on May 23, 2013 and is designated Class II. Reason for recall: This is a recall expansion from the 2010 recall which now includes seven additional lots that were not previously included. Zimmer, Inc. is initiating a recall of small and large Fukuda-type… The recall status is terminated (terminated Dec 18, 2014). Affected quantity: 722 units (includes 2010 market withdrawal).

Recall Details

Product Type
Devices
Report Date
Jul 10, 2013
Initiation Date
May 23, 2013
Termination Date
Dec 18, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
722 units (includes 2010 market withdrawal)

Product Description

Bigliani/Flatow¿ The Complete Solution Fukuda Retractors Small Nonsterile 00-4305-019-30. Used in retracting the humeral shaft posteriorly and exposing the entire glenoid surface.

Reason for Recall

This is a recall expansion from the 2010 recall which now includes seven additional lots that were not previously included. Zimmer, Inc. is initiating a recall of small and large Fukuda-type retractors due to fractures occurring at the grooves around the perimeter of the blades while being used during surgery.

Distribution Pattern

Worldwide Distribution-USA (nationwide) including the states of NY, MI, MN, OH, IN, IL, VA, MD, PA, WA, KY, CT, RI, NH, TX, MT, and CA and the countries of Chile, China, South Korea, and Australia.

Code Information

Item Number 00-4305-019-30 with Lot Numbers 61557761, 61591732, 61612275, 61614764, and 77000186. 2010 Market Withdrawal part numbers: Item #:430501930, Lot#s: 60795817, 60834923, 60916795, 60997103, 61017041, 61038697, 61075510, 61128521, 61186400, 61199809, 61253702, 61392455, 61430126

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1621-2013
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.