Z-1628-2016 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- May 18, 2016
- Initiation Date
- April 21, 2016
- Termination Date
- November 22, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 630 units
Product Description
GEM(TM), FLOWCOUPLER(R), 2.5 mm, STERILE EO, Rx only, 1 FLOWCOUPLER, REF GEM2753-FC. Product Usage: The FlowCOUPLER Device is a single use, implantable device that is intended to be used in the end-to-end anastomosis of veins and arteries normally encountered in microsurgical and vascular reconstructive procedures. The FlowCOUPLER Device includes a pair of permanently implanted rings which secure the anastomosis and a removable Doppler probe that is press-fit onto one of the rings. When the FlowCOUPLER Device is used in conjunction with the FlowCOUPLER Monitor, the FlowCOUPLER System is intended to detect blood flow and confirm vessel patency intra-operatively and post-operatively at the anastomotic site.
Reason for Recall
Instructions for use booklet may puncture the outer Tyvek lid.
Distribution Pattern
Nationwide and Germany, Sweden, Netherlands, Israel, Denmark, United Kingdom, Canada
Code Information
SP15H05-1067842 thru SP15KI9-1099303