Z-0670-2017 Class II Terminated

Recalled by Baxter Healthcare Corp. — Deerfield, IL

FDA device recall Z-0670-2017 was initiated by Baxter Healthcare Corp. on Sep 28, 2016 and is designated Class II. Reason for recall: A correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library in order to replace speakers installed in pumps or included in spare part rear case assemblies that have an… The recall status is terminated (terminated Jun 15, 2018). Affected quantity: 5,976 units (US: 5972, Canada: 4).

Recall details

Product Type
Devices
Report Date
Dec 7, 2016
Initiation Date
Sep 28, 2016
Termination Date
Jun 15, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5,976 units (US: 5972, Canada: 4)

Product description

sigma spectrum VOLUMETRIC INFUSION PUMP with Master Drug Library, V6 Spectrum Infusion Pump, Product Code 35700BAX and 35700BAXR, Unique Device Identifier GTIN 00085412091570, Rx ONLY -- Sigma, LLC Medina, NY 14103 --- NOTE: The speakers are a part of the rear case assembly located in the back of the infusion pump. The speaker is intended to generate audible tones during each keystroke, and during alarm conditions. However, some speakers have a variation in their internal wiring that increases the potential for the speaker to fail prematurely. Spectrum Infusion Pumps with non-functional speakers will continue to display visual alarm indications, and will otherwise continue to operate as intended.

Reason for recall

A correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library in order to replace speakers installed in pumps or included in spare part rear case assemblies that have an increased potential to fail prematurely.

Distribution pattern

Nationwide Distribution, Puerto Rico and Canada

Code information

Contact CDRH for list of affected serial numbers (5976 units)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0670-2017
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Baxter Healthcare Corp.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Baxter Healthcare Corp.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Baxter Healthcare Corp. →

Source: openFDA record for Z-0670-2017. Report a data issue: hello@regdatalab.com.