Z-1629-2022 Class II Ongoing

Recalled by Mckesson Medical-Surgical Inc. Corporate Office — Richmond, VA

Recall Details

Product Type
Devices
Report Date
August 31, 2022
Initiation Date
May 25, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
a. 11 b. 13 c. 15 d. 33

Product Description

BECKMAN COULTER, INC. a. COULTER DxH Diff Pack REAGENT, DIFF PK FP DXHCOULTR, Catalog #628020. b. COULTER DxH Cleaner REAGENT, CLEANER FP DXH 10LCOULTR Catalog #628023. c. DxH 500 Series Lyse LYSE, VERSA SOL DXH 500 SERIES(0.5L/KT) Catalog # B36846. d. DxH 500 Series Cleaner CLEANER, COULTER CLENZ DXH 500SERIES (0.5L/KT) Catalog # B36868

Reason for Recall

Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.

Distribution Pattern

US Nationwide

Code Information

GTIN: a. 15099590233716. b. 15099590233747. c. 15099590671860. d. 15099590671877