Z-1639-2023 Class II Ongoing

Recalled by Maquet Medical Systems USA — Wayne, NJ

FDA device recall Z-1639-2023 was initiated by Maquet Medical Systems USA on May 18, 2023 and is designated Class II. Reason for recall: 1. An accessory may be improperly placed during packaging, causing creases and damage to the accessory's sterile pouch. 2. Product packaging may exhibit holes, cracks, dents, and crushed areas… The recall status is ongoing. Affected quantity: 0 (US).

Recall details

Product Type
Devices
Report Date
Jun 7, 2023
Initiation Date
May 18, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
0 (US)

Product description

BEQ-HMO 71000-USA /QUADROX-i Adult, Oxygenator used during cardiopulmonary bypass Item no. 701067829

Reason for recall

1. An accessory may be improperly placed during packaging, causing creases and damage to the accessory's sterile pouch. 2. Product packaging may exhibit holes, cracks, dents, and crushed areas. Both issues may compromise the sterile barrier of the packaging, and exposure to non-sterile product may result in inflammation, infection, sepsis, and ischemia.

Distribution pattern

Domestic distribution nationwide.

Code information

UDI-DI: 4058863017372 All lots