Z-1648-2022 Class II Ongoing

Recalled by Siemens Healthcare Diagnostics, Inc. — Newark, DE

FDA device recall Z-1648-2022 was initiated by Siemens Healthcare Diagnostics, Inc. on Jul 18, 2022 and is designated Class II. Reason for recall: Negative bias with patient samples. The mean patient sample bias range from -26% to -33%,potential risk of delay in appropriate treatment and management decisions. The recall status is ongoing. Affected quantity: 129 units US; 159 units OUS.

Recall Details

Product Type
Devices
Report Date
Sep 7, 2022
Initiation Date
Jul 18, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
129 units US; 159 units OUS

Product Description

Dimension Vista LOCI CA 15-3 Flex reagent cartridge. In Vitro Diagnostic for the quantitative measurement of CA 15-3. Siemens Material Number (SMN)/Ref Catalog Number : 10464344/K6456

Reason for Recall

Negative bias with patient samples. The mean patient sample bias range from -26% to -33%,potential risk of delay in appropriate treatment and management decisions.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, Austria, France, ¿¿ Germany, Italy, Switzerland.

Code Information

UDI: 0084276802852621340BB22120610464344840, 0084276802852622049BC23021810464344840, 0084276802852622112BF23042210464344840. Lot Numbers: lot 21340BB Exp Date: 2022-12-06; lot 22049BC (corrected lot number 9/8/2022) Exp Date: 2023-02-18; lot 22112BF Exp Date: 2023-04-20.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1648-2022
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.