Z-1661-2023 Class II Ongoing

Recalled by Maquet Medical Systems USA — Wayne, NJ

Recall Details

Product Type
Devices
Report Date
June 7, 2023
Initiation Date
May 18, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
0 (US)

Product Description

Venous Hardshell Cardiotomy Reservoir, used during cardiopulmonary bypass Model Item No. VKMO 10000-USA 701070441 VKMO 11000-USA 701070445

Reason for Recall

Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.

Distribution Pattern

Nationwide domestic distribution

Code Information

Model Item No. UDI VKMO 10000-USA 701070441 4058863153889 VKMO 11000-USA 701070445 4058863153889 All lots