Z-1662-2023 Class II Ongoing

Recalled by Maquet Medical Systems USA — Wayne, NJ

Recall Details

Product Type
Devices
Report Date
June 7, 2023
Initiation Date
May 18, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
0 (US)

Product Description

QUADROX-iR, used during cardiopulmonary bypass Model Item No. Description BEQ-HMO 51100-USA 701067936 Small Adult with filter, with BIOLINE coating BEQ-HMO 71100-USA 701067880 Adult with filter, with BIOLINE Coating HMO 50100-USA 701067934 Small Adult without filter, with SOFTLINE Coating HMO 51100-USA 701067938 Small Adult with filter, with SOFTLINE Coating HMO 70100-USA 701067874 Adult without filter, with SOFTLINE Coating HMO 71100-USA 701067886 Adult with filter, with SOFTLINE Coating

Reason for Recall

Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.

Distribution Pattern

Nationwide domestic distribution

Code Information

Model Item No. UDI-DI BEQ-HMO 51100-USA 701067936 N/A BEQ-HMO 71100-USA 701067880 4058863164052 HMO 50100-USA 701067934 N/A HMO 51100-USA 701067938 N/A HMO 70100-USA 701067874 N/A HMO 71100-USA 701067886 N/A All lots