Z-1665-2018 Class II Terminated

Recalled by Baxter Healthcare Corporation — Medina, NY

FDA device recall Z-1665-2018 was initiated by Baxter Healthcare Corporation on Apr 3, 2018 and is designated Class II. Reason for recall: Service technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitted. he omitted tests… The recall status is terminated (terminated Feb 14, 2020). Affected quantity: 444 units in total.

Recall Details

Product Type
Devices
Report Date
May 16, 2018
Initiation Date
Apr 3, 2018
Termination Date
Feb 14, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
444 units in total

Product Description

Baxter Sigma Spectrum Infusion Pump with Master Drug Library (Version 6) 35700BAXR Product Usage: The Baxter Sigma Spectrum Infusion Pump with master Drug Library is intended to be used for the controlled administration of fluids. These fluids may includes pharmaceutical drugs, blood, blood products and mixture if required patient therapy. The intended routes of administration consist of the following clinically accepted routes: intravenous, arterial, subcutaneous, intrathecal, epidural or irrigation of fluids. Spectrum infusion pump is intended to be used in conjunction with legally marketed intravenous administration sets and medications provided by user. The Spectrum and Spectrum with MAster Drug Library is intended to reduce operator interaction through automated programming thereby helping to reduce errors associated with complex device programming.

Reason for Recall

Service technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitted. he omitted tests include air-in-line detection at 400ml/hr and 800ml/hr and verification of upstream occlusion detection at 800ml/hr. Improper detection of air-in-line or upstream occlusion may result in serious adverse health consequences.

Distribution Pattern

US Nationwide & Puerto Rico

Code Information

Serial numbers 745365 750241 750330 752814 754975 757930 762405 763117 764490 768855 772652 773031 773134 773520 774413 778845 779430 779822 781545 782706 782985 783185 787001 793146 833195 865842 883498 883854 889556 889775 890230 891398 891421 891647 891689 901881 917897 917963 919594 919736 919837 921078 921819 924734 928406 937123 940345 940860 940949 941786 944619 966019 990542 996435 996823 997156 998126 1011229 1011544 1014955 1027799

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1665-2018
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.