Z-1668-2018 Class II Terminated

Recalled by Siemens Healthcare Diagnostics, Inc. — Tarrytown, NY

FDA device recall Z-1668-2018 was initiated by Siemens Healthcare Diagnostics, Inc. on Jan 2, 2018 and is designated Class II. Reason for recall: The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine… The recall status is terminated (terminated Feb 10, 2023).

Recall Details

Product Type
Devices
Report Date
May 16, 2018
Initiation Date
Jan 2, 2018
Termination Date
Feb 10, 2023
Voluntary/Mandated
Voluntary: Firm initiated

Product Description

Dimension¿ Thyroid Stimulating Hormone (TSHL) Flex¿ reagent cartridge

Reason for Recall

The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere.

Distribution Pattern

Nationwide and Canada, Mexico, and Thailand

Code Information

TSHL, radioimmunoassay, thyroid-stimulating hormone, RF612, SMN# 10464524 Lot codes: EB7276 EC7298 EA7318 EB7346 EC7346 EA8012 EA8033 EA8051 EC8080 EB8107 EB8121 EA8143 EB8143 EA8177 EB8177 EB8199 EA8223 EA8241 EA8261

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1668-2018
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.