Z-1669-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- May 16, 2018
- Initiation Date
- January 2, 2018
- Termination Date
- February 10, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- N/A
Product Description
Dimension¿ Free Thyroxine (FT4L) Flex¿ reagent cartridge
Reason for Recall
The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere.
Distribution Pattern
Nationwide and Canada, Mexico, and Thailand
Code Information
FT4L, radioimmunoassay, free thyroxime, RF610, SMN# 10464523 Lot codes: EA7276 EC7304 EC7333 EA7346 EB8010 EA8037 EA8065 EB8086 EB8108 EA8130 EA8156 EA8179 EA8205 EA8233 FD8254