Z-1669-2018 Class II Terminated

Recalled by Siemens Healthcare Diagnostics, Inc. — Tarrytown, NY

Recall Details

Product Type
Devices
Report Date
May 16, 2018
Initiation Date
January 2, 2018
Termination Date
February 10, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
N/A

Product Description

Dimension¿ Free Thyroxine (FT4L) Flex¿ reagent cartridge

Reason for Recall

The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere.

Distribution Pattern

Nationwide and Canada, Mexico, and Thailand

Code Information

FT4L, radioimmunoassay, free thyroxime, RF610, SMN# 10464523 Lot codes: EA7276 EC7304 EC7333 EA7346 EB8010 EA8037 EA8065 EB8086 EB8108 EA8130 EA8156 EA8179 EA8205 EA8233 FD8254