Z-1671-2018 Class II Terminated

Recalled by Siemens Healthcare Diagnostics, Inc. — Tarrytown, NY

Recall Details

Product Type
Devices
Report Date
May 16, 2018
Initiation Date
January 2, 2018
Termination Date
February 10, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
N/A

Product Description

Dimension Vista¿ Cardiac Troponin I (CTNI) Flex¿ reagent cartridge

Reason for Recall

Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.

Distribution Pattern

Nationwide and Canada, Mexico, and Thailand

Code Information

CTNI, immunoassay method, troponin subunit, K6421, SMN# 10445098 Lot codes: 16336BA 16347BD 16363BC 17032AD 17045BC 17054BD 17074BC 17093BD 17121BA 17145BE 17180BB 17193BD 17201BB 17223BA 17264BC