Z-1673-2018 Class II Terminated

Recalled by Siemens Healthcare Diagnostics, Inc. — Tarrytown, NY

FDA device recall Z-1673-2018 was initiated by Siemens Healthcare Diagnostics, Inc. on Jan 2, 2018 and is designated Class II. Reason for recall: The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine… The recall status is terminated (terminated Feb 10, 2023).

Recall Details

Product Type
Devices
Report Date
May 16, 2018
Initiation Date
Jan 2, 2018
Termination Date
Feb 10, 2023
Voluntary/Mandated
Voluntary: Firm initiated

Product Description

Dimension Vista¿ Thyroid Stimulating Hormone (TSH) Flex¿ reagent cartridge

Reason for Recall

The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere.

Distribution Pattern

Nationwide and Canada, Mexico, and Thailand

Code Information

TSH, radioimmunoassay, thyroid stimulating hormone, RF612, SMN# 10464524 Lot codes: 16277BD 16322BB 17005BB 17023AC 17069AA 17093BB 17122AA 17150AA 17193BA 17215AC 17268AA