Z-1673-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- May 16, 2018
- Initiation Date
- January 2, 2018
- Termination Date
- February 10, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- N/A
Product Description
Dimension Vista¿ Thyroid Stimulating Hormone (TSH) Flex¿ reagent cartridge
Reason for Recall
The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere.
Distribution Pattern
Nationwide and Canada, Mexico, and Thailand
Code Information
TSH, radioimmunoassay, thyroid stimulating hormone, RF612, SMN# 10464524 Lot codes: 16277BD 16322BB 17005BB 17023AC 17069AA 17093BB 17122AA 17150AA 17193BA 17215AC 17268AA