Z-1674-2018 Class II Terminated

Recalled by Siemens Healthcare Diagnostics, Inc. — Tarrytown, NY

FDA device recall Z-1674-2018 was initiated by Siemens Healthcare Diagnostics, Inc. on Jan 2, 2018 and is designated Class II. Reason for recall: Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur… The recall status is terminated (terminated Feb 10, 2023).

Recall Details

Product Type
Devices
Report Date
May 16, 2018
Initiation Date
Jan 2, 2018
Termination Date
Feb 10, 2023
Voluntary/Mandated
Voluntary: Firm initiated

Product Description

IMMULITE ¿ 2000/IMMULITE ¿ 2000 XPi 3g Allergy specific IgE Universal Kit

Reason for Recall

Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.

Distribution Pattern

Nationwide and Canada, Mexico, and Thailand

Code Information

3g Allergy specific IgE Universal kit, 3g Allergy specific IgE Universal kit (IMMULITE 2000/IMMULITE 2000 XPi L2KUN6 (D)) Lot codes: 3g Allergy Specific IgE Lot # 461 462 463 464 465 466 467 468 469 470 471 472 473 475 476 3g Allergy Specific IgE Lot # D463 D464 D465 D466 D467 D469 D471 D472 D476 D483

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1674-2018
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.