Z-1681-2013 Class II Terminated

Recalled by Linvatec Corp. dba ConMed Linvatec — Largo, FL

Recall Details

Product Type
Devices
Report Date
July 17, 2013
Initiation Date
May 30, 2013
Termination Date
September 4, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4,691 units, total of each

Product Description

***REF C7122***QTY 1***APEX Arthroscopy Tubing Set. Intended for use for joint irrigation and dissention during arthroscopic procedures.

Reason for Recall

Conmed Linvatec received complaints that their products C7120 APEX Arthroscopy Tubing Set, One Connection and C7122 APEX Arthroscopy Tubing Set, Two Connection, contained a scratch/grooved area on the inside diameter of the pressure fitting that may cause the APEX Irrigation (pump) Console to sense a pressure drop and increase fluid flow to the surgical site.

Distribution Pattern

Worldwide Distribution-USA (nationwide) including the states of AR, CA, FL, GA, IL, IN, KS, MO, NC, NJ, NM, NY, OH, OK, PA, TN, TX, UT, VA, and WI, and the countries of Austria; Belgium, Canada, China, Czech Republic, Finland, France, Great Britain, Italy, and Japan.

Code Information

C7122 Lots #'s: 1301224, 1301234, 1301244, 1301304, 1301314, 1302114, 1302154, 1302274, 1302284, 1303144, 1303254, 1304044, 1304234, 1304244, and 1305024