Z-1340-2013 Class II Terminated

Recalled by Linvatec Corp. dba ConMed Linvatec — Largo, FL

Recall Details

Product Type
Devices
Report Date
May 29, 2013
Initiation Date
March 18, 2013
Termination Date
November 6, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
267 units

Product Description

Ref: VP1600 HD Still Capture System ConMed Linvatec . Captures, stores and transfers images generated by a medical endoscopic camera system during surgical procedures.

Reason for Recall

The VP1600 HD Still Capture Image System is recalled due to an anomaly between the software code in the VP1600 console and the communications interface for various keyboard key strokes (Caps lock and Num Lock) which may lock up and requires the user to reboot to continue.

Distribution Pattern

Worldwide Distribution-USA (nationwide) including the Virgin Islands and the states of CA, CT, FL, GA, IN, KY, KS, LA, MD, MI, MN, MO, MS, MT, NE, NC, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, and WI, and the countries of Austria, Belgium, Canada, Denmark, Hong Kong and Zambia.

Code Information

VP1600