Z-1681-2022 Class II Ongoing
FDA device recall Z-1681-2022 was initiated by ICU Medical, Inc. on Jun 27, 2022 and is designated Class II. Reason for recall: Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and… The recall status is ongoing. Affected quantity: 454,635.
Recall Details
- Product Type
- Devices
- Report Date
- Sep 14, 2022
- Initiation Date
- Jun 27, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 454,635
Product Description
Product Description/Item: ChemoLock Bag Spike, 10 units/CL-10-10; ChemoLock BAG SPIKE w/ ADDITIVE PORT, DRY SPIKE/CL-12; APPX 1.5ml,ChemoLock BAG SPIKE WITH ADDITIVE PORT/CL-13; ChemoLock Port/CL2100; ChemoLock Port, 5 Units/CL2100-5; ChemoLock Port/CL2100T; Clave Bag Spike w ChemoLock Port/CL2150; 30" (76 cm) Appx 3.3 ml, 20 Drop Admin Set w Integrated ChemoLock Drip Chamber, Spiros w/Red Cap, Hanger/CL3011; 9 IN(23cm)APPX 2.7ml,EXT SET,ChemoLock PORT,2 CLAMPS,2 GRADUATED CONN/CL3364; Syringe Transfer Set w Clave, ChemoLock Port/CL-34; 30" (76 cm) Appx 3.6 ml, 20 Drop Admin Set w Integrated ChemoLock Port Drip Chamber, ChemoLock w Red Cap, Bag Hanger/CL3511; 30" (76 cm) Appx 3.6 ml, 20 Drop Admin Set w Integrated ChemoLock Port Drip Chamber, ChemoLock w Red Cap, Bag Hanger/CL3511T; 30" (76cm) Appx 6.3 ml, 20 Drop Admin Set w Integrated ChemoLock Port Drip Chamber, 0.2 Micron Filter, ChemoLock w Red Cap, Hanger/CL3528; 7" (18 cm) Appx .93 ml, Ext Set w/ChemoLock, Y-Connector, Rotating Luer/CL3535T; Oncology Kit w/5" (13 cm) Add-On Set w ChemoLock Additive Port, Vented Cap, ChemoLock, Spinning Spiros w Red Cap/CL3538; Graduated Connector w ChemoLock Port/CL3900; ONCOLOGY KIT, EXT,ChemoLock PORT, GRAD CONN,TRANSFER,ChemoLock,VIAL SPIKE,ChemoLock PORT/CL3927; Oncology Kit w 5" (13 cm) Add-On Set w ChemoLock w Red Cap, Vented Cap, ChemoLock Port w Bag Spike, Spiros w Red Cap/CL3946; Oncology Kit w 17" (43 cm) Bifuse Ext Set w/ChemoLock w Red Cap, ChemoLock Port, ChemoLock Vented Vial Spike, 13mm ChemoLock/CL3947; Oncology Kit w ChemoLock Port Closed Vial Spike, ChemoLock, ChemoLock Port Bag Spike/CL3952; APPX 0.5ml, ChemoLock BAG SPIKE WITH ADDITIVE PORT,ChemoLock Port/CL3955;
Reason for Recall
Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and exposure to caustic substances. Ports available as standalone connectors, or utilized as access points on vial adaptors, bag spikes, and administration sets.
Distribution Pattern
Worldwide distribution - US Nationwide distribution including in the states of LA, FL, TX, NY, NE, WI, PA, IA, ND, CA, OH, WA, GA, AL, CO, AZ, NC, TN, MI, VA, MA, DE, OR, KY, CT, OK, NJ, MN, IL, VT, AR, RI, ID, IN, UT, NV, MT, MD, KS, SC, NM, ME, MO, SD and the countries of : Saudi Arabia, Australia, United Arab Emirates, Spain, Slovenia, Japan, Canada.
Code Information
Item/UDI-DI/Lots: CL-10-10/10887709071022/5872839, 5891255; CL-12/10887709056975/5819981, 5856055, 5873186; CL-13/10887709056982/5850811; CL2100/5797867, 5816334, 5829882,5850510; CL2100-5/10887709122960/5754377; CL2100T/10887709099293/5816341, 5842229, 5850517; CL2150/10887709094670/5829900, 5850507; CL3011/10887709057019/5829885, 5842227; CL3364/10887709073408/5841261, 5850805; CL-34/10887709069777/5754437, 5829875; CL3511/10887709057026/5842231, 5861991, 5894477; CL3511T/10887709120010/5829831, 5891263; CL3528/10887709057040/5850500, 5872833, 5883793, 5894481, 5909807; CL3535T/10887709099613/5685350; CL3538/10887709071909/5935851; CL3900/10887709071411/5772697; CL3927/10887709075440/5842211; CL3946/10887709080291/5829895, 5861983; CL3947/10887709080307/5782844; CL3952/10887709081571/5782863,5816379; CL3955/10887709082509/5841209, 5850839; CL3960/10887709083278/5816328, 5842216; CL3963/10887709083827/5842214,5851775; CL4114/10887709091136/5754773, 5782864, 5816356; CL4130/10887709095530/5850515; CL4131/10887709095776/5895940; CL4136/10887709096131/5816325, 5850661; CL4143/10887709097213/5782848, 5873243, 5891266; CL4146-5/10887709097466/5872843, 5883800, 5909812; CL4153/10887709098722/5816348; CL4159/10887709121369/5881696; CL4179/10887709123578/5872687; CL-62/10887709057088/5816320,5825248; CL-70/10887709057095/5842228; CL-70-10/10887709070988/5816343, 5851774; CL-70-10T/10887709099477/5816353; CL-70S/10887709067155/5842225,5850512; CL-70S-5/10887709070995/5816326, 5851776; CL-70T/10887709099507/5816350; CL-72/10887709062440/5829832, 5849740, 5816312; CL-72T/10887709099552/5841181, 5849748; CL-76/10887709122878/5816323, 5842218; CL-80/10887709057125/5804034, 5842200, 5850499; CL-80S/10887709069593/5808063, 5816306, 5829020, 5829023, 5841982, 5845354, 5862839; CL-80S-10/10840619080577/5804045, 5816315; CL-80S-10T/10887709120539/5804051, 5816310, 5829022, 5841994, 5849630, 5850497, 5871336; CL-80S-5/10887709077116/5667054, 5829021, 5829024, 5841980, 5849474, 5849631, 5849632, 5849713, 5849714, 5849715, 5849716, 5849474; CL-80SL/10887709069630/5816317; CL-82/10887709085692/5804065; CLH-12/10887709080413/5842212