Z-1681-2022 Class II Ongoing

Recalled by ICU Medical, Inc. — San Clemente, CA

FDA device recall Z-1681-2022 was initiated by ICU Medical, Inc. on Jun 27, 2022 and is designated Class II. Reason for recall: Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and… The recall status is ongoing. Affected quantity: 454,635.

Recall Details

Product Type
Devices
Report Date
Sep 14, 2022
Initiation Date
Jun 27, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
454,635

Product Description

Product Description/Item: ChemoLock Bag Spike, 10 units/CL-10-10; ChemoLock BAG SPIKE w/ ADDITIVE PORT, DRY SPIKE/CL-12; APPX 1.5ml,ChemoLock BAG SPIKE WITH ADDITIVE PORT/CL-13; ChemoLock Port/CL2100; ChemoLock Port, 5 Units/CL2100-5; ChemoLock Port/CL2100T; Clave Bag Spike w ChemoLock Port/CL2150; 30" (76 cm) Appx 3.3 ml, 20 Drop Admin Set w Integrated ChemoLock Drip Chamber, Spiros w/Red Cap, Hanger/CL3011; 9 IN(23cm)APPX 2.7ml,EXT SET,ChemoLock PORT,2 CLAMPS,2 GRADUATED CONN/CL3364; Syringe Transfer Set w Clave, ChemoLock Port/CL-34; 30" (76 cm) Appx 3.6 ml, 20 Drop Admin Set w Integrated ChemoLock Port Drip Chamber, ChemoLock w Red Cap, Bag Hanger/CL3511; 30" (76 cm) Appx 3.6 ml, 20 Drop Admin Set w Integrated ChemoLock Port Drip Chamber, ChemoLock w Red Cap, Bag Hanger/CL3511T; 30" (76cm) Appx 6.3 ml, 20 Drop Admin Set w Integrated ChemoLock Port Drip Chamber, 0.2 Micron Filter, ChemoLock w Red Cap, Hanger/CL3528; 7" (18 cm) Appx .93 ml, Ext Set w/ChemoLock, Y-Connector, Rotating Luer/CL3535T; Oncology Kit w/5" (13 cm) Add-On Set w ChemoLock Additive Port, Vented Cap, ChemoLock, Spinning Spiros w Red Cap/CL3538; Graduated Connector w ChemoLock Port/CL3900; ONCOLOGY KIT, EXT,ChemoLock PORT, GRAD CONN,TRANSFER,ChemoLock,VIAL SPIKE,ChemoLock PORT/CL3927; Oncology Kit w 5" (13 cm) Add-On Set w ChemoLock w Red Cap, Vented Cap, ChemoLock Port w Bag Spike, Spiros w Red Cap/CL3946; Oncology Kit w 17" (43 cm) Bifuse Ext Set w/ChemoLock w Red Cap, ChemoLock Port, ChemoLock Vented Vial Spike, 13mm ChemoLock/CL3947; Oncology Kit w ChemoLock Port Closed Vial Spike, ChemoLock, ChemoLock Port Bag Spike/CL3952; APPX 0.5ml, ChemoLock BAG SPIKE WITH ADDITIVE PORT,ChemoLock Port/CL3955;

Reason for Recall

Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and exposure to caustic substances. Ports available as standalone connectors, or utilized as access points on vial adaptors, bag spikes, and administration sets.

Distribution Pattern

Worldwide distribution - US Nationwide distribution including in the states of LA, FL, TX, NY, NE, WI, PA, IA, ND, CA, OH, WA, GA, AL, CO, AZ, NC, TN, MI, VA, MA, DE, OR, KY, CT, OK, NJ, MN, IL, VT, AR, RI, ID, IN, UT, NV, MT, MD, KS, SC, NM, ME, MO, SD and the countries of : Saudi Arabia, Australia, United Arab Emirates, Spain, Slovenia, Japan, Canada.

Code Information

Item/UDI-DI/Lots: CL-10-10/10887709071022/5872839, 5891255; CL-12/10887709056975/5819981, 5856055, 5873186; CL-13/10887709056982/5850811; CL2100/5797867, 5816334, 5829882,5850510; CL2100-5/10887709122960/5754377; CL2100T/10887709099293/5816341, 5842229, 5850517; CL2150/10887709094670/5829900, 5850507; CL3011/10887709057019/5829885, 5842227; CL3364/10887709073408/5841261, 5850805; CL-34/10887709069777/5754437, 5829875; CL3511/10887709057026/5842231, 5861991, 5894477; CL3511T/10887709120010/5829831, 5891263; CL3528/10887709057040/5850500, 5872833, 5883793, 5894481, 5909807; CL3535T/10887709099613/5685350; CL3538/10887709071909/5935851; CL3900/10887709071411/5772697; CL3927/10887709075440/5842211; CL3946/10887709080291/5829895, 5861983; CL3947/10887709080307/5782844; CL3952/10887709081571/5782863,5816379; CL3955/10887709082509/5841209, 5850839; CL3960/10887709083278/5816328, 5842216; CL3963/10887709083827/5842214,5851775; CL4114/10887709091136/5754773, 5782864, 5816356; CL4130/10887709095530/5850515; CL4131/10887709095776/5895940; CL4136/10887709096131/5816325, 5850661; CL4143/10887709097213/5782848, 5873243, 5891266; CL4146-5/10887709097466/5872843, 5883800, 5909812; CL4153/10887709098722/5816348; CL4159/10887709121369/5881696; CL4179/10887709123578/5872687; CL-62/10887709057088/5816320,5825248; CL-70/10887709057095/5842228; CL-70-10/10887709070988/5816343, 5851774; CL-70-10T/10887709099477/5816353; CL-70S/10887709067155/5842225,5850512; CL-70S-5/10887709070995/5816326, 5851776; CL-70T/10887709099507/5816350; CL-72/10887709062440/5829832, 5849740, 5816312; CL-72T/10887709099552/5841181, 5849748; CL-76/10887709122878/5816323, 5842218; CL-80/10887709057125/5804034, 5842200, 5850499; CL-80S/10887709069593/5808063, 5816306, 5829020, 5829023, 5841982, 5845354, 5862839; CL-80S-10/10840619080577/5804045, 5816315; CL-80S-10T/10887709120539/5804051, 5816310, 5829022, 5841994, 5849630, 5850497, 5871336; CL-80S-5/10887709077116/5667054, 5829021, 5829024, 5841980, 5849474, 5849631, 5849632, 5849713, 5849714, 5849715, 5849716, 5849474; CL-80SL/10887709069630/5816317; CL-82/10887709085692/5804065; CLH-12/10887709080413/5842212

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1681-2022
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.