Z-1682-2022 Class II Ongoing

Recalled by ICU Medical, Inc. — San Clemente, CA

Recall Details

Product Type
Devices
Report Date
September 14, 2022
Initiation Date
June 27, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
358,975

Product Description

Product Description/Item: ChemoLock Vial Spike, 20mm/011-CL-80S; ChemoLock Bag Spike/CL-10; ChemoLock w/Mini Bag Spike/CL-17; 8 IN (20cm) APPX 1.0ml, EXT SET, MicroClave Clear, ChemoLock PORT, YCONN/CL3950; 31" (79 cm) Appx 3.4 ml, Admin Set w/20 Drop Integrated ChemoLock Drip Chamber, Spiros, Purple Cap, Hanger/CL3951; 7.5" (19 cm) Appx 1.5 ml, Ext Set w/ChemoLock Port, Clamp, Graduated Adapter/CL3967; 6 IN(15cm)APPX 0.82ml,EXT SET,MicroClave Clear,ChemoLock PORT,YCONN, ROTATING LUER/CL4118; ChemoLock Kit w/ 5" (13 cm) Bag Spike Adapter w/ChemoLock w/Red Cap, Vented Cap, ChemoLock Port Bag Spike/CL4158; ChemosafeLock Bag Spike/KL-BS001U3; ChemosafeLock Vial Adapter/KL-VA201U3; ChemosafeLock Vial Adapter/KL-VA202U3; ChemosafeLock Vial Adapter/KL-VA321U3; ChemosafeLock Vial Adapter/KL-VA322U3; ChemoLock Vented Vial Spike/011-CL-51; Transfer Set, PUR Yellow w/ChemoLock Port, Spiros/011-CL4155; Puerto ChemoLock/034-CL2100; ChemoLock, Punzon Univ. Vial c/ Filtro, fijacion/034-CL-70; ChemoLock, Punzon Vial Tritan c/ Filtro, 13mm/034-CL-72

Reason for Recall

Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and exposure to caustic substances. Ports available as standalone connectors, or utilized as access points on vial adaptors, bag spikes, and administration sets.

Distribution Pattern

Worldwide distribution - US Nationwide distribution including in the states of LA, FL, TX, NY, NE, WI, PA, IA, ND, CA, OH, WA, GA, AL, CO, AZ, NC, TN, MI, VA, MA, DE, OR, KY, CT, OK, NJ, MN, IL, VT, AR, RI, ID, IN, UT, NV, MT, MD, KS, SC, NM, ME, MO, SD and the countries of : Saudi Arabia, Australia, United Arab Emirates, Spain, Slovenia, Japan, Canada.

Code Information

OUS Distributed Only Item/Lots: 011-CL-80S/5843950; CL-10/5850516; CL-17/5816314, 5829878, 5842209, 5850503; CL3950/5845117, 5885020; CL3951/5842199, 5872831; CL3967/5829901, 5842210; CL4118/5825274, 5872840; KL-BS001U3/5851020, 5851021, 5851022, 5910255, 5910258; KL-VA201U3/5819364; KL-VA202U3/5819370, 5910262; KL-VA321U3/5915973; KL-VA322U3/5920744; 011-CL-51/5850505; 011-CL4155/5816324; 034-CL2100/5842230; 034-CL-70/5816321; 034-CL-72/5816344