Z-1682-2022 Class II Ongoing
FDA device recall Z-1682-2022 was initiated by ICU Medical, Inc. on Jun 27, 2022 and is designated Class II. Reason for recall: Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and… The recall status is ongoing. Affected quantity: 358,975.
Recall Details
- Product Type
- Devices
- Report Date
- Sep 14, 2022
- Initiation Date
- Jun 27, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 358,975
Product Description
Product Description/Item: ChemoLock Vial Spike, 20mm/011-CL-80S; ChemoLock Bag Spike/CL-10; ChemoLock w/Mini Bag Spike/CL-17; 8 IN (20cm) APPX 1.0ml, EXT SET, MicroClave Clear, ChemoLock PORT, YCONN/CL3950; 31" (79 cm) Appx 3.4 ml, Admin Set w/20 Drop Integrated ChemoLock Drip Chamber, Spiros, Purple Cap, Hanger/CL3951; 7.5" (19 cm) Appx 1.5 ml, Ext Set w/ChemoLock Port, Clamp, Graduated Adapter/CL3967; 6 IN(15cm)APPX 0.82ml,EXT SET,MicroClave Clear,ChemoLock PORT,YCONN, ROTATING LUER/CL4118; ChemoLock Kit w/ 5" (13 cm) Bag Spike Adapter w/ChemoLock w/Red Cap, Vented Cap, ChemoLock Port Bag Spike/CL4158; ChemosafeLock Bag Spike/KL-BS001U3; ChemosafeLock Vial Adapter/KL-VA201U3; ChemosafeLock Vial Adapter/KL-VA202U3; ChemosafeLock Vial Adapter/KL-VA321U3; ChemosafeLock Vial Adapter/KL-VA322U3; ChemoLock Vented Vial Spike/011-CL-51; Transfer Set, PUR Yellow w/ChemoLock Port, Spiros/011-CL4155; Puerto ChemoLock/034-CL2100; ChemoLock, Punzon Univ. Vial c/ Filtro, fijacion/034-CL-70; ChemoLock, Punzon Vial Tritan c/ Filtro, 13mm/034-CL-72
Reason for Recall
Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and exposure to caustic substances. Ports available as standalone connectors, or utilized as access points on vial adaptors, bag spikes, and administration sets.
Distribution Pattern
Worldwide distribution - US Nationwide distribution including in the states of LA, FL, TX, NY, NE, WI, PA, IA, ND, CA, OH, WA, GA, AL, CO, AZ, NC, TN, MI, VA, MA, DE, OR, KY, CT, OK, NJ, MN, IL, VT, AR, RI, ID, IN, UT, NV, MT, MD, KS, SC, NM, ME, MO, SD and the countries of : Saudi Arabia, Australia, United Arab Emirates, Spain, Slovenia, Japan, Canada.
Code Information
OUS Distributed Only Item/Lots: 011-CL-80S/5843950; CL-10/5850516; CL-17/5816314, 5829878, 5842209, 5850503; CL3950/5845117, 5885020; CL3951/5842199, 5872831; CL3967/5829901, 5842210; CL4118/5825274, 5872840; KL-BS001U3/5851020, 5851021, 5851022, 5910255, 5910258; KL-VA201U3/5819364; KL-VA202U3/5819370, 5910262; KL-VA321U3/5915973; KL-VA322U3/5920744; 011-CL-51/5850505; 011-CL4155/5816324; 034-CL2100/5842230; 034-CL-70/5816321; 034-CL-72/5816344
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1682-2022
- openFDA file date
- Oct 23, 2026
- Loaded by RegDataLab