Z-1687-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- May 16, 2018
- Initiation Date
- January 2, 2018
- Termination Date
- February 10, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- N/A
Product Description
ADVIA Centaur¿ Cyclosporine
Reason for Recall
The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere
Distribution Pattern
Nationwide and Canada, Mexico, and Thailand
Code Information
ADVIA Centaur¿ Cyclosporine; Cyclosporine; SMN 10335448 Lot codes: ADVIA Centaur CSA Lot # 38182025 64432025 98554025 58299030 25547030 30153031 77934031 97170031