Z-1687-2018 Class II Terminated

Recalled by Siemens Healthcare Diagnostics, Inc. — Tarrytown, NY

Recall Details

Product Type
Devices
Report Date
May 16, 2018
Initiation Date
January 2, 2018
Termination Date
February 10, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
N/A

Product Description

ADVIA Centaur¿ Cyclosporine

Reason for Recall

The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere

Distribution Pattern

Nationwide and Canada, Mexico, and Thailand

Code Information

ADVIA Centaur¿ Cyclosporine; Cyclosporine; SMN 10335448 Lot codes: ADVIA Centaur CSA Lot # 38182025 64432025 98554025 58299030 25547030 30153031 77934031 97170031