Z-1688-2018 Class II Terminated

Recalled by Siemens Healthcare Diagnostics, Inc. — Tarrytown, NY

FDA device recall Z-1688-2018 was initiated by Siemens Healthcare Diagnostics, Inc. on Jan 2, 2018 and is designated Class II. Reason for recall: The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine… The recall status is terminated (terminated Feb 10, 2023).

Recall Details

Product Type
Devices
Report Date
May 16, 2018
Initiation Date
Jan 2, 2018
Termination Date
Feb 10, 2023
Voluntary/Mandated
Voluntary: Firm initiated

Product Description

ADVIA Centaur¿ DHEA-SO4

Reason for Recall

The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere

Distribution Pattern

Nationwide and Canada, Mexico, and Thailand

Code Information

ADVIA Centaur¿ DHEA-SO4; radioimmunoassay, dehydroepiandrosterone (free and sulfate); SMN 10282227 ADVIA Centaur DHEA-S Kit Lot # 00119053 14041053 80449053 90520053 29750054 30423054 39155054 50533054 63267055 68936055 79273055 86427055 06402057 07109057 14187057 37499057

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1688-2018