Z-1694-2015 Class II Terminated

Recalled by Zimmer, Inc. — Warsaw, IN

FDA device recall Z-1694-2015 was initiated by Zimmer, Inc. on Apr 29, 2015 and is designated Class II. Reason for recall: The entire scope of 4.5mm Cortical Screws listed as compatible with the M/DN system is not compatible with the M/DN system. This has the potential to lead to a situation in which the screw would bind… The recall status is terminated (terminated Dec 8, 2015). Affected quantity: 289,616.

Recall Details

Product Type
Devices
Report Date
Jun 10, 2015
Initiation Date
Apr 29, 2015
Termination Date
Dec 8, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
289,616

Product Description

Cortical Screw for M/DN Intramedullary Fixation. Metaphyseal/Diaphyseal Nail System, Orthopedic surgical device.

Reason for Recall

The entire scope of 4.5mm Cortical Screws listed as compatible with the M/DN system is not compatible with the M/DN system. This has the potential to lead to a situation in which the screw would bind upon entering the nail and have to be removed during the surgery or removal of the screw after fracture healing.

Distribution Pattern

Worldwide Distribution - US (nationwide) and Internationally to Canada, Brazil, Mexico, Bahamas, Nicaragua, Singapore, China, Japan, Malaysia, Taiwan, Thailand, Korea, Spain, Belgium, Switzerland, India, Germany, England, Italy, Saudi Arabia, Lebanon, Libya, United Arab Emirates, Algeria, France, United Kingdom, Canary Islands, Iraq, Jordan, Lithuania, Morocco, Macedonia, Malta, Mauritius, Serbia, Russian Federation, and Yemen.

Code Information

All lots. Part numbers: 00-2253-020-45 INTERLOCKING IM SCREW 4.5MM DIA X 20MM 00-2253-022-45 INTERLOCKING IM SCREW 4.5MM DIA X 22.5MM 00-2253-025-45 INTERLOCKING IM SCREW 4.5MM DIA X 25MM 00-2253-027-45 M/DN Intramedullary Fixation 4.5mm Dia. Cortical Screw 00-2253-030-45 INTERLOCKING IM SCREW 4.5MM DIA X 30MM 00-2253-032-45 INTERLOCKING IM SCREW 4.5MM DIA X 32.5MM 00-2253-035-45 INTERLOCKING IM SCREW 4.5MM DIA X 35MM 00-2253-037-45 INTERLOCKING IM SCREW 4.5MM DIA X 37.5MM 00-2253-040-45 INTERLOCKING IM SCREW 4.5MM DIA X 40MM 00-2253-042-45 INTERLOCKING IM SCREW 4.5MM DIA X 42.5MM 00-2253-045-45 INTERLOCKING IM SCREW 4.5MM DIA X 45MM 00-2253-047-45 INTERLOCKING IM SCREW 4.5MM DIA X 47.5MM 00-2253-050-45 INTERLOCKING IM SCREW 4.5MM DIA X 50MM 00-2253-052-45 INTERLOCKING IM SCREW 4.5MM DIA X 52.5MM 00-2253-055-45 INTERLOCKING IM SCREW 4.5MM DIA X 55MM 00-2253-057-45 INTERLOCKING IM SCREW 4.5MM DIA X 57.5MM 00-2253-060-45 INTERLOCKING IM SCREW 4.5MM DIA X 60MM 00-2253-062-45 INTERLOCKING IM SCREW 4.5MM DIA X 62.5MM 00-2253-065-45 INTERLOCKING IM SCREW 4.5MM DIA X 65MM 00-2253-067-45 INTERLOCKING IM SCREW 4.5MM DIA X 67.5MM 00-2253-070-45 INTERLOCKING IM SCREW 4.5MM DIA X 70MM 00-2253-072-45 INTERLOCKING IM SCREW 4.5MM DIA X 72.5MM 00-2253-075-45 INTERLOCKING IM SCREW 4.5MM DIA X 75MM 00-2253-077-45 INTERLOCKING IM SCREW 4.5MM DIA X 77.5MM 00-2253-080-45 INTERLOCKING IM SCREW 4.5MM DIA X 80MM 00-2253-082-45 INTERLOCKING IM SCREW 4.5MM DIA X 82.5MM 00-2253-085-45 INTERLOCKING IM SCREW 4.5MM DIA X 85MM 00-2253-087-45 INTERLOCKING IM SCREW 4.5MM DIA X 87.5MM 00-2253-090-45 INTERLOCKING IM SCREW 4.5MM DIA X 90MM

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1694-2015
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.