Z-1695-2023 Class III Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- June 14, 2023
- Initiation Date
- April 14, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 10752 units
Product Description
RR Mechatronics Control Level A ESR Control4.6 ml QRR049002- In-vitro diagnostic quality control material to monitor the precision of Starrsed Erythrocyte Sedimentation Rate (ESR) instruments and procedures. Product Code: A0026578
Reason for Recall
QRR049001 Lot QCAB6AN505 incorrectly labelled as QRR049002 Lot QCAB6AAA28. The incorrectly labelled vial will upload the range for the Starrsed Control Level A and the instrument result will be flagged as out of range.
Distribution Pattern
Worldwide distribution - US Nationwide distribution in the state of OH and the country of Netherlands.
Code Information
UDI: (01)08719189137125(17)240222(10)QCAB6AAA28 Lot Number/Exp. Date: QCAB6AAA28 24-02-2022