Z-1694-2023 Class III Ongoing

Recalled by Mechatronics USA — Warwick, RI

Recall Details

Product Type
Devices
Report Date
June 14, 2023
Initiation Date
April 14, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
9306 units

Product Description

RR Mechatronics Starrsed Control Level N ESR Control 4.6 ml QRR049001- In-vitro diagnostic quality control material to monitor the precision of Starrsed Erythrocyte Sedimentation Rate (ESR) instruments and procedures. Product Code: A0026577

Reason for Recall

QRR049001 Lot QCAB6AN505 incorrectly labelled as QRR049002 Lot QCAB6AAA28. The incorrectly labelled vial will upload the range for the Starrsed Control Level A and the instrument result will be flagged as out of range.

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the state of OH and the country of Netherlands.

Code Information

UDI-DI: (01)08719189137118(17)240222(10)QCAB6AN505 Lot Number/Exp Date: QCAB6AN505 24-02-2022