Z-1714-2016 Class II Terminated

Recalled by Philips Healthcare — Andover, MA

Recall Details

Product Type
Devices
Report Date
June 15, 2016
Initiation Date
September 6, 2015
Termination Date
March 4, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
US-423

Product Description

Philips Healthcare Brilliance iCT Computed Tomography X-Ray System

Reason for Recall

Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of view; (4) DoseRight algorithm miscalculations leading to incorrect CTDI values; (5) scan length changes during subsequent axial results, (6) surview scan lengths near 135mm or 184mm

Distribution Pattern

US Distribution - AZ, PA, FL, DC, W VA, NC, SC, AR, OK, OH, NY, NV, TX, PA, IN, CA, MN, MA, DC, MS, MD, OR, IL, CO, MO, VT, IA, AL, LA, GA and WA.

Code Information

Software v. 4.13, 4.14 and 4.15.