Z-1719-2026 Class I Ongoing

Recalled by Medline Industries, LP — Northfield, IL

Recall Details

Product Type
Devices
Report Date
April 15, 2026
Initiation Date
February 27, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2980 kits

Product Description

Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. NON VASCULAR PACK, Medline SKU # DYNJ68037C; 2. VEIN, Medline SKU # DYNJ89845; 3. VEIN, Medline SKU # DYNJ89845A.

Reason for Recall

Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Distribution Pattern

Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).

Code Information

Medline SKU # DYNJ68037C, UDI/DI each 10198459239328, UDI/DI case 40198459239329, Lot Number: 25HBV397; Medline SKU # DYNJ89845, UDI/DI each 10198459292460, UDI/DI case 40198459292461, Lot Number: 25CBU222; Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 26BBC195; Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 26ABI611; Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 25LBK402; Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 25KBH371; Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 25IBK975; Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 25HBQ426.