Z-1727-2026 Class I Ongoing

Recalled by Medline Industries, LP — Northfield, IL

Recall Details

Product Type
Devices
Report Date
April 15, 2026
Initiation Date
February 27, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
966 kits

Product Description

Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. DELIVERY PACK, Medline SKU # DYNJ42892F; 2. DELIVERY PACK, Medline SKU # DYNJ42892G.

Reason for Recall

Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Distribution Pattern

Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).

Code Information

Medline SKU # DYNJ42892F, UDI/DI each10195327693718, UDI/DI case 40195327693719, Lot Number : 25GMD696; Medline SKU # DYNJ42892F, UDI/DI each10195327693718, UDI/DI case 40195327693719, Lot Number : 25EMJ573; Medline SKU # DYNJ42892F, UDI/DI each10195327693718, UDI/DI case 40195327693719, Lot Number : 25DMC646; Medline SKU # DYNJ42892F, UDI/DI each10195327693718, UDI/DI case 40195327693719, Lot Number : 25BMD630; Medline SKU # DYNJ42892F, UDI/DI each10195327693718, UDI/DI case 40195327693719, Lot Number : 24LMA523; Medline SKU # DYNJ42892F, UDI/DI each10195327693718, UDI/DI case 40195327693719, Lot Number : 24KMA084; Medline SKU # DYNJ42892F, UDI/DI each10195327693718, UDI/DI case 40195327693719, Lot Number : 24JME306; Medline SKU # DYNJ42892F, UDI/DI each10195327693718, UDI/DI case 40195327693719, Lot Number : 24FME624; Medline SKU # DYNJ42892F, UDI/DI each10195327693718, UDI/DI case 40195327693719, Lot Number : 24EMF099; Medline SKU # DYNJ42892G, UDI/DI each10198459504570, UDI/DI case 40198459504571, Lot Number : 26BMH149; Medline SKU # DYNJ42892G, UDI/DI each10198459504570, UDI/DI case 40198459504571, Lot Number : 26AMD029; Medline SKU # DYNJ42892G, UDI/DI each10198459504570, UDI/DI case 40198459504571, Lot Number : 25KMA060; Medline SKU # DYNJ42892G, UDI/DI each10198459504570, UDI/DI case 40198459504571, Lot Number : 25IMA514.