Z-1734-2015 Class II Terminated

Recalled by Roche Diabetes Care, Inc. — Indianapolis, IN

FDA device recall Z-1734-2015 was initiated by Roche Diabetes Care, Inc. on Apr 22, 2015 and is designated Class II. Reason for recall: Roche has become aware of an increase of mechanical errors experienced by customers using the Accu-Chek Spirit Combo insulin infusion pump and Accu-Chek Spirit insulin infusion pump when the insulin… The recall status is terminated (terminated Nov 3, 2016). Affected quantity: 35,795.

Recall Details

Product Type
Devices
Report Date
Jun 17, 2015
Initiation Date
Apr 22, 2015
Termination Date
Nov 3, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
35,795

Product Description

AccuChek Spirit insulin infusion pump.

Reason for Recall

Roche has become aware of an increase of mechanical errors experienced by customers using the Accu-Chek Spirit Combo insulin infusion pump and Accu-Chek Spirit insulin infusion pump when the insulin pumps display E6 and El0 error messages. If the user does not act upon the E6 and E10 error messages appropriately, insulin delivery will be stopped and if unnoticed may lead to severe hyperglycemia u

Distribution Pattern

Nationwide Distribution

Code Information

All Serial Numbers Models: 1. 05458544001, Accu-Chek Combo Kit mg US/English version. 2. 05504686001, Accu-Chek Combo Kit mg US/Spanish version. 3. 05391512001, Accu-Chek Spirit Combo Replacement Pump US/EN. 4. 05391750001, Accu-Chek Spirit Combo Demo Pump  US. 5. 04759702001, Accu-Chek Spirit Pump Kit US/CA Back Up. 6. 04540492001, Accu-Chek Spirit Pump Kit US/CA. 7. 04759729001, Accu-Chek Spirit Replacement Pump.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1734-2015