Z-1734-2026 Class II Ongoing
FDA device recall Z-1734-2026 was initiated by Civco Medical Instruments Co. Inc. on Mar 2, 2026 and is designated Class II. Reason for recall: There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user… The recall status is ongoing. Affected quantity: 20 units.
Recall details
- Product Type
- Devices
- Report Date
- Apr 8, 2026
- Initiation Date
- Mar 2, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 20 units
Product description
Reason for recall
Distribution pattern
US States: OH, PA, WA. China.
Code information
Source
- Dataset