Z-1892-2025 Class II Ongoing

Recalled by Civco Medical Instruments Co. Inc. — Kalona, IA

FDA device recall Z-1892-2025 was initiated by Civco Medical Instruments Co. Inc. on May 1, 2025 and is designated Class II. Reason for recall: During the attachment of the needle guide to the covered probe, the tight tolerance and fit of the needle guide can cause a pinhole breach in the transducer cover when latching the guide in place. The recall status is ongoing. Affected quantity: 506 boxes.

Recall details

Product Type
Devices
Report Date
Jun 11, 2025
Initiation Date
May 1, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
506 boxes

Product description

BX2 Needle Guide, Part Number 644-094, tool for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers

Reason for recall

During the attachment of the needle guide to the covered probe, the tight tolerance and fit of the needle guide can cause a pinhole breach in the transducer cover when latching the guide in place.

Distribution pattern

OUS Distribution to countries of: Japan, Taiwan, France, Georgia, Germany and Denmark

Code information

UDI-DI: 00841436120050; Lot Numbers: A174424, A176855, A177208, A182744, A184023, A185336, A185337, A186904, A189734, A190489, A193623, A194755, A198716, A198795, A199940, A200430, A202208, A203169, A203201, A203520, A205457, A207901, A208740, A211619, A212291, A213027, A215531, A216885, A217412, A220102, A221365, A221925, A225145, A226800, A228446, A228778, A229092, A229898, A230045, A230054, A230549, A230581, A231101, A231810, A232903, A233075, A233535, A236538, A238767, A239339, A239355, A240873, A240931, A241445, A242018, A242161, A243263, A243264, A245342, A245527, A246173, A248520, A248572, A250325, A252513, A252842, A254080, A254081, A254900, A254955, A255332, A257534, A258150, A258926, A260607

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1892-2025
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Civco Medical Instruments Co. Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1892-2025. Report a data issue: hello@regdatalab.com.