Z-1892-2025 Class II Ongoing

Recalled by Civco Medical Instruments Co. Inc. — Kalona, IA

Recall Details

Product Type
Devices
Report Date
June 11, 2025
Initiation Date
May 1, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
506 boxes

Product Description

BX2 Needle Guide, Part Number 644-094, tool for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers

Reason for Recall

During the attachment of the needle guide to the covered probe, the tight tolerance and fit of the needle guide can cause a pinhole breach in the transducer cover when latching the guide in place.

Distribution Pattern

OUS Distribution to countries of: Japan, Taiwan, France, Georgia, Germany and Denmark

Code Information

UDI-DI: 00841436120050; Lot Numbers: A174424, A176855, A177208, A182744, A184023, A185336, A185337, A186904, A189734, A190489, A193623, A194755, A198716, A198795, A199940, A200430, A202208, A203169, A203201, A203520, A205457, A207901, A208740, A211619, A212291, A213027, A215531, A216885, A217412, A220102, A221365, A221925, A225145, A226800, A228446, A228778, A229092, A229898, A230045, A230054, A230549, A230581, A231101, A231810, A232903, A233075, A233535, A236538, A238767, A239339, A239355, A240873, A240931, A241445, A242018, A242161, A243263, A243264, A245342, A245527, A246173, A248520, A248572, A250325, A252513, A252842, A254080, A254081, A254900, A254955, A255332, A257534, A258150, A258926, A260607