Z-1741-2015 Class II Terminated

Recalled by Biomet, Inc. — Warsaw, IN

FDA device recall Z-1741-2015 was initiated by Biomet, Inc. on May 1, 2015 and is designated Class II. Reason for recall: Ankle clamp was assembled incorrectly; the offset of the pin to the pivot point is reversed. This prevents complete range of motion and if pushed too far may loosen the pin from the bone. A revision… The recall status is terminated (terminated Dec 8, 2015). Affected quantity: 19.

Recall Details

Product Type
Devices
Report Date
Jun 17, 2015
Initiation Date
May 1, 2015
Termination Date
Dec 8, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
19

Product Description

Unilateral external fixation ankle clamp Product Usage: Unilateral external fixation for use treatment of bone conditions amenable to treatment by use of external fixation modality

Reason for Recall

Ankle clamp was assembled incorrectly; the offset of the pin to the pivot point is reversed. This prevents complete range of motion and if pushed too far may loosen the pin from the bone. A revision surgery may be necessary to adequately reconnect the fixator to the bone.

Distribution Pattern

US Nationwide Distribution in the states of CA, TX, FL, NC, MI, GA.

Code Information

Part 14-450510 Lots:268000, 466170, and 813950