Z-1751-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- May 16, 2018
- Initiation Date
- March 29, 2018
- Termination Date
- April 9, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 220
Product Description
Connecting Tube - Female to Male Luer Lock, HPCT8.8-60-M-FM Product Usage: The connecting tube is intended to be a conduit between a fluid reservoir and a vascular catheter in the patient. The connecting tube may be used during angiographic procedures to connect radiographic contrast power injector to a vascular catheter, arterial pressure monitoring, and fluid infusion.
Reason for Recall
Cap-adapter assembly of the device may potentially not be adequately tightened, possibly leading to fluid leakage from or air aspiration into the connecting tube.
Distribution Pattern
Distributed to accounts in 7 states: AL, CA, GA, NE, NV, NY, and TX. International distribution to Japan, Korea, and Taiwan.
Code Information
lots 7853883 7871665 7940332