Z-1752-2020 Class II Terminated

Recalled by Beckman Coulter Inc. — Brea, CA

Recall Details

Product Type
Devices
Report Date
April 29, 2020
Initiation Date
September 4, 2019
Termination Date
April 28, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7,951 units

Product Description

Opiate 300 Ng Reagent (OP 1 X 250) for use in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems OP 300 Urine Calibrators, Catalogue No.475024. The OP assay provides a rapid screening procedure for determining the presence of OP and its metabolites in urine.

Reason for Recall

Cross-reactivity testing may not meet the claims indicated in the Opiate 300 ng (OP)Instructions For Use. The presence of certain compounds in urine may result in false positive or false negative patient test results.

Distribution Pattern

Distributed throughout the United States and to Puerto Rico. States include: CA, NC, OR, MI, GE, AL, SC, TX, NJ, IA, MS, IN, MA, NC, OH, AZ, LA, MA, NY, PA, KY, WV, MI, WA, CO, IL, VA, FL, KS, MD, ID, DE, NV, OK, AR, HI, WI, NY.

Code Information

Catalog Number 475024; UDI:15099590222970 ; Lot Numbers: M904130 and older including M802097, M803045, M804024, M805123, M806080, M808069, M809218, M810293, M901044, M902086, and M903175. .