Z-1754-2018 Class II Terminated

Recalled by Cook Inc. — Bloomington, IN

FDA device recall Z-1754-2018 was initiated by Cook Inc. on Mar 29, 2018 and is designated Class II. Reason for recall: Cap-adapter assembly of the device may potentially not be adequately tightened, possibly leading to fluid leakage from or air aspiration into the connecting tube. The recall status is terminated (terminated Apr 9, 2019). Affected quantity: 304.

Recall Details

Product Type
Devices
Report Date
May 16, 2018
Initiation Date
Mar 29, 2018
Termination Date
Apr 9, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
304

Product Description

Connecting Tube - Female to Male Luer Lock, HPCT8.8-80-M-FM Product Usage: The connecting tube is intended to be a conduit between a fluid reservoir and a vascular catheter in the patient. The connecting tube may be used during angiographic procedures to connect radiographic contrast power injector to a vascular catheter, arterial pressure monitoring, and fluid infusion.

Reason for Recall

Cap-adapter assembly of the device may potentially not be adequately tightened, possibly leading to fluid leakage from or air aspiration into the connecting tube.

Distribution Pattern

Distributed to accounts in 7 states: AL, CA, GA, NE, NV, NY, and TX. International distribution to Japan, Korea, and Taiwan.

Code Information

Lots 7844043, 7898128, 7940347, NS7904210

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report