Z-1765-2020 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

Recall Details

Product Type
Devices
Report Date
May 13, 2020
Initiation Date
April 6, 2020
Termination Date
October 27, 2020
Voluntary/Mandated
FDA Mandated
Product Quantity
399 (wordwide)

Product Description

System AXIOM Aristos FX Plus - A dedicated x-ray system with a flat panel detector which allows the acquisition of x-ray exposures without the use of conventional film/screen systems; and allows radiographic exposures of the whole body including skull, spinal column, chest, abdomen, and extremities. It is not designed for mammography use. Radiographic exposures may be taken with the patient in the sitting, standing, or supine positions.

Reason for Recall

It was determined that in rare cases persons can receive low dose of unintended radiation. The system performed exposures without the exposure button being pressed. Exposure will be released for less than 500ms, therefore, the overall dose is estimated at 0.01mSv or less than 500 mGy in total.

Distribution Pattern

Worldwide Distribution

Code Information

serail number 1192