Z-1775-2022 Class II Ongoing

Recalled by Covidien Llc — Mansfield, MA

FDA device recall Z-1775-2022 was initiated by Covidien Llc on Aug 19, 2022 and is designated Class II. Reason for recall: Medtronic has confirmed six reports from customers outside the US where the ventilator became inoperable due to loss of gas supply caused by manufacturing error of specific turbine components. The recall status is ongoing. Affected quantity: 5 US.

Recall Details

Product Type
Devices
Report Date
Sep 28, 2022
Initiation Date
Aug 19, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5 US

Product Description

Puritan Bennett 560 Ventilator, PB560, CFN 4096600

Reason for Recall

Medtronic has confirmed six reports from customers outside the US where the ventilator became inoperable due to loss of gas supply caused by manufacturing error of specific turbine components.

Distribution Pattern

Domestic distribution to one customer in Washington, D.C. Foreign distribution worldwide.

Code Information

UDI-DI: 10884521196377 U.S. Serial Numbers: 40966N7229 40966N7231 40966N7334 40966N7335 40966N7340