Z-1779-2024 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- May 15, 2024
- Initiation Date
- March 18, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1 unit
Product Description
GE Healthcare Interact Discovery RT, Interventional fluoroscopic x-ray systems
Reason for Recall
GE HealthCare has become aware of the potential that adhesive material might not have been applied to some bolts of certain Innova IGS, Discovery IGS, Interact IGS, Allia IGS and Allia Pulse IGS products. This could result in a potential fall of the Detector/Elevator or X-ray tube of the frontal Gantry assembly.
Distribution Pattern
Worldwide distribution - US Nationwide in the states of CA, FL, GA, IN, MA, NC, OK, SC, TN, TX, VA and the countries of Armenia, China, France, Hungary, India, Italy, Japan, Kazakhstan, Lithuania, Mexico, Russia, Spain, Taiwan, Turkey, Vietnam
Code Information
UDI/DI 00840682141581, Serial Numbers: M4-23-049