Z-1783-2024 Class II Ongoing

Recalled by GE Medical Systems, SCS — Buc, N/A

FDA device recall Z-1783-2024 was initiated by GE Medical Systems, SCS on Mar 18, 2024 and is designated Class II. Reason for recall: GE HealthCare has become aware of the potential that adhesive material might not have been applied to some bolts of certain Innova IGS, Discovery IGS, Interact IGS, Allia IGS and Allia Pulse IGS… The recall status is ongoing. Affected quantity: 1 unit.

Recall details

Product Type
Devices
Report Date
May 15, 2024
Initiation Date
Mar 18, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit

Product description

GE Healthcare Allia IGS 5 Pulse, Interventional fluoroscopic x-ray systems

Reason for recall

GE HealthCare has become aware of the potential that adhesive material might not have been applied to some bolts of certain Innova IGS, Discovery IGS, Interact IGS, Allia IGS and Allia Pulse IGS products. This could result in a potential fall of the Detector/Elevator or X-ray tube of the frontal Gantry assembly.

Distribution pattern

Worldwide distribution - US Nationwide in the states of CA, FL, GA, IN, MA, NC, OK, SC, TN, TX, VA and the countries of Armenia, China, France, Hungary, India, Italy, Japan, Kazakhstan, Lithuania, Mexico, Russia, Spain, Taiwan, Turkey, Vietnam

Code information

UDI/DI 00195278719263, Serial Numbers: M3-23-117

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1783-2024
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1783-2024. Report a data issue: hello@regdatalab.com.