Z-1791-2023 Class I Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- June 28, 2023
- Initiation Date
- May 10, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 71 units
Product Description
ICD-VR VISIA AF XT OUS IS1/DF1, Model Number DVAB2D1; Implantable Cardioverter Defibrillators
Reason for Recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
Distribution Pattern
US Nationwide - Worldwide Distribution
Code Information
GTIN 00643169720756, Lot Serial Numbers: BWR600821S, BWR600947S, BWR600936S, BWR600937S, BWR600941S, BWR600946S, BWR600959S, BWR601046S, BWR600938S, BWR601096S, BWR601093S, BWR600979S, BWR601092S, BWR600997S, BWR600934S, BWR600965S, BWR601050S, BWR601049S, BWR600823S, BWR601053S, BWR600983S, BWR600989S, BWR600984S, BWR601055S, BWR601057S, BWR601098S, BWR600822S, BWR600980S, BWR600981S, BWR600982S, BWR601097S, BWR601100S, BWR601095S, BWR601101S, BWR600991S, BWR601051S, BWR601058S, BWR601099S, BWR601094S; GTIN BWR601149S, BWR601155S, BWR601137S, BWR601138S, BWR601108S, BWR601113S, BWR601111S, BWR601119S, BWR601144S, BWR601146S, BWR601147S, BWR601148S, BWR601067S, BWR601150S, BWR601152S, BWR601115S, BWR601105S, BWR601112S, BWR601059S, BWR601061S, BWR601062S, BWR601063S, BWR601072S, BWR601073S, BWR601078S, BWR601081S, BWR601090S, BWR601116S, BWR601103S, BWR601075S, BWR601102S, BWR601104S