Z-1796-2023 Class I Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- June 28, 2023
- Initiation Date
- May 10, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4 units
Product Description
ICD-VR DVBB1D4 EVERA XT DF4 US, Model Number DVBB1D4; Implantable Cardioverter Defibrillators
Reason for Recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
Distribution Pattern
US Nationwide - Worldwide Distribution
Code Information
GTIN 00643169720435, Lot Serial Number: BWH224619H, BWH224621H, BWH224620H, BWH224623H