Z-1794-2022 Class III Ongoing

Recalled by Mckesson Medical-Surgical Inc. Corporate Office — Richmond, VA

FDA device recall Z-1794-2022 was initiated by Mckesson Medical-Surgical Inc. Corporate Office on May 25, 2022 and is designated Class III. Reason for recall: Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21. The recall status is ongoing. Affected quantity: a. 12 bags, b. 4 bags.

Recall Details

Product Type
Devices
Report Date
Oct 5, 2022
Initiation Date
May 25, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
a. 12 bags, b. 4 bags

Product Description

a. Comply INDICATOR, COMPLY VAPOR GAS (250/BG 4BG/CS) 3M Model Number: 1248. b. 3M Attest INDICATOR, CHEMICAL VAPORIZEDHYDROGEN PEROXIDE (250/BG) Model Number: 1348.

Reason for Recall

Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.

Distribution Pattern

US Nationwide distribution.

Code Information

GTIN: a. 00707387508296; 00707387457037; 50707387508291; 30707387508297; 30707387457038. b. 50707387788006; 30707387788002; 00707387788001.