Z-1798-2015 Class II Terminated

Recalled by Biomet, Inc. — Warsaw, IN

FDA device recall Z-1798-2015 was initiated by Biomet, Inc. on Apr 22, 2015 and is designated Class II. Reason for recall: The Left and Right orientation of the trial was incorrectly etched on one side. When the instrument is rotated about the axis of the handle 180 degrees to view the other side, the etch on the handle… The recall status is terminated (terminated Mar 22, 2016). Affected quantity: 27.

Recall Details

Product Type
Devices
Report Date
Jun 24, 2015
Initiation Date
Apr 22, 2015
Termination Date
Mar 22, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
27

Product Description

Vanguard M Unicondylar Tibial Trial For use during knee joint replacement.

Reason for Recall

The Left and Right orientation of the trial was incorrectly etched on one side. When the instrument is rotated about the axis of the handle 180 degrees to view the other side, the etch on the handle now shows R on the same side as the keel marked LM and the L on the same side as the keel marked RM. This is incorrect. Please note that the keel is not etched on this side by design.

Distribution Pattern

Worldwide Distribution -- US, including the state of IL, NJ, Canada, Netherlands, and Chile.

Code Information

Part 32-423101 Lot 414930 Part 32-423102 Lot 116490 Part 32-423106 Lot 785220 Part 32-423107 Lot 873780 Part 32-423108 Lot 070420 and 785250 Part 32-423109 Lot 670630 Part 32-423115 Lot 672920 Part 32-423119 Lot 785290 Part 32-423116 Lot 073640 Part 32-423101 Lot 785170 Part 32-423102 Lot 785180 Part 32-423109 Lot 785260 Part 32-423124 Lot 024760 Part 32-423110 Lot 785270

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1798-2015
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.