Z-1831-2016 Class II Terminated

Recalled by Smiths Medical ASD, Inc. — Saint Paul, MN

FDA device recall Z-1831-2016 was initiated by Smiths Medical ASD, Inc. on Apr 22, 2016 and is designated Class II. Reason for recall: Smiths Medical became aware that under delivery of medication may occur on CADD Administration Sets with Flow Stop. The recall status is terminated (terminated Dec 13, 2017). Affected quantity: 13,696,830 units.

Recall Details

Product Type
Devices
Report Date
Jun 8, 2016
Initiation Date
Apr 22, 2016
Termination Date
Dec 13, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13,696,830 units

Product Description

CADD Administration Sets with Flow Stop, Product Reorder No: 21-7321-01, 21-7321-24, 21-7322-01, 21-7322-24, 21-7323-24, 21-7324-01, 21-7324-24, 21-7333-24, 21-7336-01, 21-7336-24, 21-7339-01, 21-7339-24, 21-7359-01, 21-7359-24, 21-7383-01, 21-7383-24, 21-7390-01, 21-7390-24, 21-7391-01, 21-7391-24, 21-7394-01, 21-7394-24, 21-7395-24. The CADD Medication Cassette Reservoirs with Flow Stop are a modification to the current CADD Medication Cassette Reservoirs. The reservoirs will incorporate a set-based free flow protection component (i.e. Flow Stop) that is designed to occlude the tube if the reservoirs are accidentally placed onto the pump incorrectly or become detached from the pump. The Flow Stop will be located on the reservoir housing, which is attached to the pump. The reservoir will be provided to the user in an open state. Before the reservoir can be attached to the pump, the blue "CLIP" must be removed to activate the Flow Stop. However, after attaching the reservoir to the pump, the user can still remove the reservoir from the pump and prime it by holding the Flow Stop in the open position.

Reason for Recall

Smiths Medical became aware that under delivery of medication may occur on CADD Administration Sets with Flow Stop.

Distribution Pattern

US: ALL STATES IN CONTINENTAL USA INCLUDING DC AND PR EXCEPT DE. OUS: UNITED ARAB EMIRATES, AUSTRIA BELGIUM, CANADA, CHINA, GERMANY, DENMARK, SPAIN, FINLAND, FRANCE, GREAT BRITAIN, HONG KONG, IRELAND, INDIA, ITALY, JAPAN, LUXEMBOURG, MAURITIUS, NEW CALEDONIA, NEW ZEALAND, PANAMA, PORTUGAL, SAUDI ARABIA, SWEDEN, SINGAPORE, TAIWAN, VIET NAAM, SOUTH AFRICA.

Code Information

Devices with an expiration date on or before March 2021

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1831-2016
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.